Which side of the impaction substrate for the SKC Deployable Particulate Sampler (DPS) System should be used for optimum particle capture?

These impaction substrates have a smooth, shiny side and a rough, dull side. For optimum particle capture efficiency, place substrate…

These impaction substrates have a smooth, shiny side and a rough, dull side. For optimum particle capture efficiency, place substrate smooth, shiny-side down in the impaction plate.

Which filters can be used with the Deployable Particulate System (DPS)?

The DPS System containing the Leland Legacy Sample Pump, is most often used with the following filters: Part Number 225-1823…

The DPS System containing the Leland Legacy Sample Pump, is most often used with the following filters: Part Number 225-1823 (47 mm quartz) and Part Number 225-1747 (47 mm PTFE with PMP support ring). The quartz filter is used for PM mass; the PTFE filter is used for elemental analysis.